Ad-weighing ML-50 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Ad-weighing ML-50. A&D Weighing ML-50 User Manual [en] Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1

ふ MS-70 MX-50 MF-50 ML-50 WM+PD4000477D

Page 2

MS-70 / MX-50 / MF-50 / ML-50 8 4.4.4.4. Outline and Outline and Outline and Outline and FeaturesFeaturesFeaturesFeatures The moisture an

Page 3 - ContentsContents

MS-70 / MX-50 / MF-50 / ML-50 9 The analyzer is equipped with a serial interface as standard. It can be connected to a printer or computer.

Page 4

MS-70 / MX-50 / MF-50 / ML-50 10 5.5.5.5. Packing List and Names of each partPacking List and Names of each partPacking List and Names of each pa

Page 5

MS-70 / MX-50 / MF-50 / ML-50 11 All AccessoriesAll AccessoriesAll AccessoriesAll Accessories Caution Please confirm tha

Page 6

MS-70 / MX-50 / MF-50 / ML-50 12 5.1.5.1.5.1.5.1. Display and KeysDisplay and KeysDisplay and KeysDisplay and Keys Name State and Me

Page 7

MS-70 / MX-50 / MF-50 / ML-50 13 #1 ML-50 can select "standard drying " and "quick drying" only. #2 ML-50 does not display &quo

Page 8

MS-70 / MX-50 / MF-50 / ML-50 14 6.6.6.6. PreparationPreparationPreparationPreparationssss 6.1.6.1.6.1.6.1. Installing the AnalyzerInsta

Page 9

MS-70 / MX-50 / MF-50 / ML-50 15 6.2.6.2.6.2.6.2. Setting the Clock and CSetting the Clock and CSetting the Clock and CSetting the Clock and Calend

Page 10

MS-70 / MX-50 / MF-50 / ML-50 16 6.3.6.3.6.3.6.3. Proper Operation for Precision MeasurementProper Operation for Precision MeasurementProper Oper

Page 11

MS-70 / MX-50 / MF-50 / ML-50 17 6.3.3.6.3.3.6.3.3.6.3.3. The Glass Fiber SheetThe Glass Fiber SheetThe Glass Fiber SheetThe Glass Fiber Sheetssss

Page 12

1.1.1.1. This manual and MarksThis manual and MarksThis manual and MarksThis manual and Marks All safety messages are identified by the fol

Page 13

MS-70 / MX-50 / MF-50 / ML-50 18 7.7.7.7. Measurement ProcedureMeasurement ProcedureMeasurement ProcedureMeasurement Procedure 7.1.7.1.7.1.7.

Page 14 - 5td, quc, U-a, U-t, U-m

MS-70 / MX-50 / MF-50 / ML-50 19 1. Turn on the analyzer. The gram unit (of weighing mode) is displayed. Enter the Standard ModeEnter the Stan

Page 15

MS-70 / MX-50 / MF-50 / ML-50 20 11. Open the heater cover. Put in a sample using the level indicator. Caution The sample needs to be m

Page 16

MS-70 / MX-50 / MF-50 / ML-50 21 7.2.7.2.7.2.7.2. Quick Mode OperationQuick Mode OperationQuick Mode OperationQuick Mode Operation The quic

Page 17 - Cl adj

MS-70 / MX-50 / MF-50 / ML-50 22 1. Turn on the analyzer. The gram unit (of weighing mode) is displayed. Enter the Quick ModeEnter the Quick

Page 18 - 0.000g

MS-70 / MX-50 / MF-50 / ML-50 23 11. Open the heater cover. Put in a sample using the level indicator. Caution The sample needs to be mor

Page 19

MS-70 / MX-50 / MF-50 / ML-50 24 7.3.7.3.7.3.7.3. Program NumberProgram NumberProgram NumberProgram Number The measurement conditions of

Page 20

MS-70 / MX-50 / MF-50 / ML-50 25 8.8.8.8. MeasurementMeasurementMeasurementMeasurement Programs Programs Programs Programs 8.1.8.1.8.1.8.1. Li

Page 21

MS-70 / MX-50 / MF-50 / ML-50 26 8.1.1.8.1.1.8.1.1.8.1.1. ACCURACY of the Standard Mode and Quick ModeACCURACY of the Standard Mode and Quick Mode

Page 22

MS-70 / MX-50 / MF-50 / ML-50 27 8.1.2.8.1.2.8.1.2.8.1.2. Analyzing mode of the Automatic ModeAnalyzing mode of the Automatic ModeAnalyzing mode of

Page 23

MS-70 / MX-50 / MF-50 / ML-50 1 ContentsContentsContentsContents 1. Safety and Compliance...

Page 24

MS-70 / MX-50 / MF-50 / ML-50 28 How to Select quick mode (Extracts)How to Select quick mode (Extracts)How to Select quick mode (Extracts)How to Se

Page 25

MS-70 / MX-50 / MF-50 / ML-50 29 8.2.8.2.8.2.8.2. Procedures to Store a Measurement ProgramProcedures to Store a Measurement ProgramProcedures to S

Page 26

MS-70 / MX-50 / MF-50 / ML-50 30 Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SEL

Page 27 - STOP key. Max

MS-70 / MX-50 / MF-50 / ML-50 31 Set Set Set Set the Minimum Scale Value of the Gram Displaythe Minimum Scale Value of the Gram Displaythe Minimum S

Page 28 - HI , MID. or LO

MS-70 / MX-50 / MF-50 / ML-50 32 8.2.2.8.2.2.8.2.2.8.2.2. Ramp DryingRamp DryingRamp DryingRamp Drying Standard Standard Standard Standard Mode

Page 29

MS-70 / MX-50 / MF-50 / ML-50 33 Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SELEC

Page 30 - quc using the ↓ or ↑ key

MS-70 / MX-50 / MF-50 / ML-50 34 Store Parameters and Finish the OperationStore Parameters and Finish the OperationStore Parameters and Finish the

Page 31

MS-70 / MX-50 / MF-50 / ML-50 35 Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SELEC

Page 32 - U-a

MS-70 / MX-50 / MF-50 / ML-50 36 8.2.3.8.2.3.8.2.3.8.2.3. Step DryingStep DryingStep DryingStep Drying Standard Mode, Automatic Mode or Standar

Page 33

MS-70 / MX-50 / MF-50 / ML-50 37 Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SELEC

Page 34

MS-70 / MX-50 / MF-50 / ML-50 2 8. Check Function...41 8.1. Self-Check Fu

Page 35

MS-70 / MX-50 / MF-50 / ML-50 38 Set the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % DisplaySet the Minimum Scale Val

Page 36

MS-70 / MX-50 / MF-50 / ML-50 39 1. Display gram unit (of the weighing mode). Select a Program Number to Edit the Measurement Select a Progra

Page 37

MS-70 / MX-50 / MF-50 / ML-50 40 Set tSet tSet tSet the he he he Measurement Measurement Measurement Measurement UnitUnitUnitUnit 11. Select

Page 38

MS-70 / MX-50 / MF-50 / ML-50 41 9.9.9.9. Check FunctionCheck FunctionCheck FunctionCheck Function 9.1.9.1.9.1.9.1. SelfSelfSelfSelf----Check

Page 39

MS-70 / MX-50 / MF-50 / ML-50 42 9.2.9.2.9.2.9.2. Test Sample (STest Sample (STest Sample (STest Sample (Sodium odium odium odium TTTTartrate art

Page 40 - "0."0"1%

MS-70 / MX-50 / MF-50 / ML-50 43 10.10.10.10. Connecting to a PrinterConnecting to a PrinterConnecting to a PrinterConnecting to a Printer

Page 41 - U-t of the timer mode with

MS-70 / MX-50 / MF-50 / ML-50 44 10.1.10.1.10.1.10.1. Print samplesPrint samplesPrint samplesPrint samples 10.1.1.10.1.1.10.1.1.10.1.1. Exampl

Page 42 - "0"0"1%

MS-70 / MX-50 / MF-50 / ML-50 45 10.1.2.10.1.2.10.1.2.10.1.2. Example To Print Selected ItemsExample To Print Selected ItemsExample To Print Selecte

Page 43 - CH pa55

MS-70 / MX-50 / MF-50 / ML-50 46 Print ExamplePrint ExamplePrint ExamplePrint Example üüüüüüüüüüüAü&üD Factory MODELüüüüüüMX-50 Product S/

Page 44 - 15.80%

MS-70 / MX-50 / MF-50 / ML-50 47 10.1.3.10.1.3.10.1.3.10.1.3. ExplExplExplExplanation for Print Itemanation for Print Itemanation for Print Itemanat

Page 45

MS-70 / MX-50 / MF-50 / ML-50 3 2.2.2.2. Safety Safety Safety Safety and Complianceand Complianceand Complianceand Compliance WARNING Do no

Page 46 - Signature. Refer to 10.1.3

MS-70 / MX-50 / MF-50 / ML-50 48 """"Measurement Measurement Measurement Measurement Unit"Unit"Unit"Unit"

Page 47

MS-70 / MX-50 / MF-50 / ML-50 49 11.11.11.11. Connecting to a ComputerConnecting to a ComputerConnecting to a ComputerConnecting to a Computer

Page 48

MS-70 / MX-50 / MF-50 / ML-50 50 11.1.11.1.11.1.11.1. RSRSRSRS----232C Serial Interface232C Serial Interface232C Serial Interface232C Serial Inter

Page 49

MS-70 / MX-50 / MF-50 / ML-50 51 11.2.11.2.11.2.11.2. Output FormatOutput FormatOutput FormatOutput Format In Case of Format omitted Temperature

Page 50

MS-70 / MX-50 / MF-50 / ML-50 52 In Case of Format included Temperature Data (Function Table In Case of Format included Temperature Data (Function

Page 51

MS-70 / MX-50 / MF-50 / ML-50 53 12.12.12.12. Data Memory FunctionData Memory FunctionData Memory FunctionData Memory Function The data mem

Page 52

MS-70 / MX-50 / MF-50 / ML-50 54 12.1.2.12.1.2.12.1.2.12.1.2. Output All Data at One TimeOutput All Data at One TimeOutput All Data at One TimeOut

Page 53

MS-70 / MX-50 / MF-50 / ML-50 55 13.13.13.13. CalibrationCalibrationCalibrationCalibration The moisture content is calculated with a ratio

Page 54

MS-70 / MX-50 / MF-50 / ML-50 56 13.2.13.2.13.2.13.2. CalibCalibCalibCalibration of the Weighting Sensorration of the Weighting Sensorration of th

Page 55 - data "1

MS-70 / MX-50 / MF-50 / ML-50 57 Calibration RCalibration RCalibration RCalibration Reporteporteporteport Example for the Weighi Example for the Weig

Page 56

MS-70 / MX-50 / MF-50 / ML-50 4 Compliance with FCC RulesCompliance with FCC RulesCompliance with FCC RulesCompliance with FCC Rules Please

Page 57 - Cl adj

MS-70 / MX-50 / MF-50 / ML-50 58 13.3.13.3.13.3.13.3. Calibration of Drying TemperatureCalibration of Drying TemperatureCalibration of Drying Temp

Page 58

MS-70 / MX-50 / MF-50 / ML-50 59 10. After fifteen minutes, the buzzer sounds and the display blinks "1"6"0°C . Adjust the bl

Page 59 - } Calibration type

MS-70 / MX-50 / MF-50 / ML-50 60 14.14.14.14. Function TableFunction TableFunction TableFunction Table The function table can store the f

Page 60 - 100℃ "9"7℃

MS-70 / MX-50 / MF-50 / ML-50 61 14.1.1.14.1.1.14.1.1.14.1.1. OperationOperationOperationOperation 1. Display the gram unit (of the weighing m

Page 61

MS-70 / MX-50 / MF-50 / ML-50 62 15.15.15.15. MaintenanceMaintenanceMaintenanceMaintenance Turn off the power switch and remove power cor

Page 62 - ENTER key

MS-70 / MX-50 / MF-50 / ML-50 63 15.2.15.2.15.2.15.2. Replacement of the Halogen LampReplacement of the Halogen LampReplacement of the Halogen LampR

Page 63 - dp "1

MS-70 / MX-50 / MF-50 / ML-50 64 15.3.15.3.15.3.15.3. Factory SettingsFactory SettingsFactory SettingsFactory Settings This function can s

Page 64

MS-70 / MX-50 / MF-50 / ML-50 65 15.4.15.4.15.4.15.4. TroubleshootingTroubleshootingTroubleshootingTroubleshooting 1111....In the Case thatIn th

Page 65

MS-70 / MX-50 / MF-50 / ML-50 66 15.5.15.5.15.5.15.5. Error MessageError MessageError MessageError Message CH no Internal Error An inter

Page 66

MS-70 / MX-50 / MF-50 / ML-50 67 16.16.16.16. SpecificationsSpecificationsSpecificationsSpecifications MS-70 MX-50 MF-50 ML-50 Measurement met

Page 67

MS-70 / MX-50 / MF-50 / ML-50 5 3.3.3.3. PrecPrecPrecPrecautionsautionsautionsautions 3.1.3.1.3.1.3.1. Installing the AnalyzerInstalling the A

Page 68

MS-70 / MX-50 / MF-50 / ML-50 68 MS-70 MX-50 MF-50 ML-50 Operation environment 5°C to 40°C (41°F to 104°F), 85%RH or less (no condensation) Sample

Page 69

MS-70 / MX-50 / MF-50 / ML-50 69 16.2.16.2.16.2.16.2. Accessories and Accessories and Accessories and Accessories and PeripheralPeripheralPeripheral

Page 70

MS-70 / MX-50 / MF-50 / ML-50 70 17.17.17.17. IndexIndexIndexIndex %/min ---------------------------- 18, 21, 26, 28 , ENTER key -----------

Page 71

MS-70 / MX-50 / MF-50 / ML-50 71 ISO ------------------------------- 43, 56, 59, 60 Key mode-------------------------------------- 60 Keys-----------

Page 72

MS-70 / MX-50 / MF-50 / ML-50 72 MEMO

Page 73

MS-70 / MX-50 / MF-50 / ML-50 73

Page 74

MS-70 / MX-50 / MF-50 / ML-50 6 3.2.3.2.3.2.3.2. During UseDuring UseDuring UseDuring Use ------------Caution for MCaution for MCaution for M

Page 75

MS-70 / MX-50 / MF-50 / ML-50 7 Caution for Heating (Drying). When the drying temperature is set to 200°C and measurement is started, the the