Ad-weighing SV-100 Manuel d'utilisateur Page 47

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 66
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 46
45
SV-100 output format example
Viscosity
/
Temper-
ature
Display Output format example Remarks
L Pas
LAB-12,2003/03/19,12:34:56,+025.67,C,+00000.00,
Pa s
Zeroes are output for
below measuring
range error.
1.00 Pas
LAB-12,2003/03/19,12:34:56,+025.67,C,+00001.00,
Pa s
10.0 Pas
LAB-12,2003/03/19,12:34:56,+025.67,C,+00010.00,
Pa s
Pas
/
°C
H Pas
LAB-12,2003/03/19,12:34:56,+025.67,C,+00120.00,
Pa s
120 is output for
above measuring
range error.
L P
LAB-12,2003/03/19,12:34:56,+051.23,F,+000000.0,
P
Zeroes are output for
below measuring
range error.
10.0 P
LAB-12,2003/03/19,12:34:56,+051.23,F,+000010.0,
P
100 P
LAB-12,2003/03/19,12:34:56,+051.23,F,+000100.0,
P
P
/
°F
H P
LAB-12,2003/03/19,12:34:56,+051.23,F,+001200.0,
P
1200 is output for
above measuring
range error.
: Space (ASC 20h)